Clean the instrument

Clean the instrument after each extended deployment to remove deposits that may have accumulated.

Do not use an ultrasonic bath to clean your instruments. Ultrasonic vibrations can break the wire bonding inside the transducers.

Type Procedure Notes

General/biofouling

To clean the exterior, soak in a mild detergent, then scrub the instrument with a soft brush.

Avoid scratching the plastic (scratches make future cleaning more difficult).

Calcification,
encrustation

Soak in vinegar for six hours, then scrub the surface using a soft brush.

Soaking in vinegar for more than 24 hours may damage the O-ring and increase the risk of a leak.

Cleaning the pressure sensor

Avoid touching the diaphragm when cleaning the sensor. Any deformation will permanently affect performance.

  1. Unscrew the sensor guard using a coin or a large flathead screwdriver. Do not apply excessive force, especially when using the screwdriver.

  2. Rinse the area under running water. If this fails to remove the deposits, try soaking in vinegar.

  3. If unsuccessful, contact RBR.

Cleaning ODO sensors

When dirty, carefully wipe the sensors with a soft cloth. To remove encrustation, soak in water until soft. It may take hours or days, depending on the severity.

Do not use abrasive cloths as scratched faces can affect calibration.
Do not use solvents or cleaners as these could affect optical properties of the window.

Cleaning turbidity sensors

Clean the optical face of the sensor with a lint-free cloth and isopropyl alcohol.

Ensure you are wearing gloves, oil from fingerprints easily soils the windows.