Filling the syringe and de-gassing the buffer oil
|
All required materials for this procedure are provided in the support kit. |
It is important to remove all gases from the system as they can form bubbles and cause anomalies in the data. Refill the system with de-gassed oil any time when cleaning it, or if it has had an oil leak for any reason.
Required materials
|
Recommended handling materials
|
|
Buffer oil is not a hazardous substance, but it is recommended to practice good industrial hygiene and safety practices, and to use this material in a well-ventilated space. |
Filling the syringe
| Step | Description |
|---|---|
1 |
Remove the stopper from the syringe. |
2 |
Install the needle. |
3 |
Draw 1-2ml of the oil into the syringe. |
De-gassing the buffer oil
| Step | Description | Image | |
|---|---|---|---|
1 |
Invert the syringe so that the needle is facing up and pull any remaining oil out of the needle into the syringe. |
|
|
2 |
Remove the needle. |
||
3 |
Gently push the plunger to purge the air from the syringe. |
||
4 |
Install the stopper. |
||
5 |
Reverse the syringe so that the stopper is facing down. |
||
6 |
Draw out the plunger of the syringe past the 10ml point. |
||
7 |
Install the syringe kickstand so that it cups the plunger and supports it in the drawn-out position. The syringe will brace against the flange on the plunger and the barrel. |
||
8 |
Leave the syringe in the reverse position for about an hour. |
||
9 |
Remove the kickstand. |
||
10 |
Invert the syringe so that the tip is facing up. |
||
11 |
Remove the stopper. |
||
12 |
Purge any air from the syringe |
||