Temperature

Parameter Value

Range1

-5 to 35°C

Initial accuracy

±0.002°C

Resolution

<0.00005°C

Typical stability

±0.002°C / year

Time constant

<0.1s | fast, <1s standard, <60s | slow

1 A wider temperature range is available upon request. Contact RBR for more information.

Calibration equations

Temperature

\[T_m=\ln\left(\left(\frac{1}{V}\right)-1\right)\]
\[T_c=\frac{1}{(C_0+C_1T_m+C_2T^2_m+C_3T^3_m)}-273.15\]

where:

Tm = Intermediate variables

V = Raw temperature channel voltage ratio output (unique with each individual measurement)

C0 = Cubic function regression calibration coefficient

C1 = Cubic function regression calibration coefficient

C2 = Cubic function regression calibration coefficient

C3 = Cubic function regression calibration coefficient

Tc = Temperature in celsius