Conductivity (C)
RBR CT | uv and CTD | uv instruments use integrated inductive conductivity sensors which measure the ability of seawater to conduct electric current.
| Parameter | Value |
|---|---|
Range |
0 to 85mS/cm |
Initial accuracy |
±0.003mS/cm |
Resolution |
<0.0001mS/cm |
Typical stability1 |
±0.010mS/cm/year |
Max depth rating |
200m |
1 For a clean cell with no biofouling.
The CT cell is rugged and durable. It is not affected by surface contaminants or freezing conditions, and can be frozen into ice.
While 80% of its volumetric measurements happens inside the cell, they also extend up to 15cm away and thus may be affected by conductive and non-conductive objects within this distance. RBR calibrates conductivity sensors to account for static objects, such as cages, guards, and other sensors.
|
To maintain optimal accuracy, deploy your instrument at least 15cm away from other objects. |
Conductivity measurements are temperature compensated.