Cleaning the instrument
Clean the instrument after each extended deployment to remove deposits that may have accumulated.
|
Do not use an ultrasonic bath to clean your instruments. Ultrasonic vibrations can break the wire bonding inside the transducers. |
| Type | Procedure | Notes |
|---|---|---|
General/biofouling |
To clean the exterior, soak in a mild detergent, then scrub the instrument with a soft brush. |
Avoid scratching the plastic (scratches make future cleaning more difficult). |
Calcification, |
Soak in vinegar for six hours, then scrub the surface using a soft brush. |
Soaking in vinegar for more than 24 hours may damage the O-ring and increase the risk of a leak. |
Cleaning the pressure sensor
|
Avoid touching the diaphragm when cleaning the sensor. Any deformation will permanently affect performance. |
-
Unscrew the sensor guard using a coin or a large flathead screwdriver. Do not apply excessive force, especially when using the screwdriver.
-
Rinse the area under running water. If this fails to remove the deposits, try soaking in vinegar.
-
If unsuccessful, contact RBR.
Cleaning the CT cell
|
Typically, the white ceramic tube inside the CT cell does not need cleaning because the UV radiation protects it from biofouling. If some cleaning is still required, use a very thin cloth. Feed it throw the tube and floss back and forth. |
-
Feed the cloth through the guard and remove the biofouling from all surfaces.
-
Rinse the area under running water.
-
If unsuccessful, contact RBR.
Cleaning the housing
Scrub the surface using a soft brush.