RBRcoda³ D
The RBRcoda³ D and RBRcoda³ D | deep (Fig. 2) use the piezoresistive pressure sensors.
The sensor is protected by a clear plastic guard. During deployments, always orient the sensor downwards to reduce debris collecting in the housing.
Pressure
| Parameter | Value |
|---|---|
Range |
20 / 50 / 100 / 200 / 500 / 1000dbar (plastic) |
Initial accuracy |
±0.05% full scale |
Resolution |
<0.001% full scale |
Typical stability |
±0.05% full scale / year |
Time constant |
<10ms |
Physical
| Parameter | Value |
|---|---|
Connector |
MCBH-6-MP |
Communications |
RS-232 |
Housing |
Plastic or titanium |
Diameter |
25mm |
Length |
~200mm (~235mm with connector) |
Weight |
170g in air, 70g in water (plastic) |
Depth rating |
Up to 1000m (plastic) |
Sampling rate |
2Hz standard, 8Hz | fast8, 16Hz | fast16, | tide16 |
Power
| Parameter | Value |
|---|---|
Supply voltage |
6V to 18V (12V nominal) |
Power |
10mA (120mW) |
Calibration equations
Raw pressure |
where:
\(C_0-C_3\)are unit-specific calibration coefficients determined in the laboratory.
\(VR\) is the measured Voltage Ratio
Temperature compensated pressure
where:
\(P_{raw}\) is the uncorrected pressure
\(P_{cal}\)is the nominal pressure while performing the temperature calibration segment
(stored as "X0" in the instrument)
\(T_{pres}\) is the temperature of the pressure sensor
\(Tcal\) is the average temperature during the room-temperature pressure calibration run
(stored as "X5" in the instrument).
\(X_1-X_4\) are unit-specific calibration coefficients determined in the laboratory