RBRquartz³ BPR with Ethernet
At the time of order, you can request your RBRquartz³ BPR wired for Ethernet and power. See RBRquartz³ BPR interface for more details on end-cap types.
|
You do not need to open the instrument to connect it to your network. Follow the steps below. |
Deploying the RBRquartz³ BPR
-
Connect your terminal to the MINK or MCBH port located outside the instrument body using a compatible patch cable. The instrument will appear on your network.
-
Connect to the control port (2323).
-
Enter the question mark (“?”) command, terminated with linefeed, to display the network controller control menu.
------ RBRquartz3 BPR NETWORK INTERFACE CONTROL MENU ------D: Display current settings S: Set NTP source R: Set NTP refresh rate J: Set the timestamp jump tolerance E: Set the transparent mode inactivity timeout T: Enter transparent mode !: Exit transparent mode (once in transparent mode) Q: Close this connection #: Reset the RBRquartz3 BPR network interface ?: Print this menu -
Enter “T” to enable the transparent mode.
-
Enter “clock” to reset the instrument clock if necessary.
-
Enter deployment parameters, for example:
deployment starttime = 20000101000000, endtime = 20991231235959 sampling mode = continuous, period = 63 -
Erase the memory and start the deployment.
enable erasememory = true
|
For more information on deploying the RBRquartz³ BPR with Ethernet, see Ethernet commands. |