RBRquartz³ BPR with Ethernet

At the time of order, you can request your RBRquartz³ BPR wired for Ethernet and power. See RBRquartz³ BPR interface for more details on end-cap types.

You do not need to open the instrument to connect it to your network. Follow the steps below.

Deploying the RBRquartz³ BPR

  1. Connect your terminal to the MINK or MCBH port located outside the instrument body using a compatible patch cable. The instrument will appear on your network.

  2. Connect to the control port (2323).

  3. Enter the question mark (“?”) command, terminated with linefeed, to display the network controller control menu.

    ------ RBRquartz3 BPR NETWORK INTERFACE CONTROL MENU ------D: Display current settings
    S: Set NTP source
    R: Set NTP refresh rate
    J: Set the timestamp jump tolerance
    E: Set the transparent mode inactivity timeout
    T: Enter transparent mode
    !: Exit transparent mode (once in transparent mode)
    Q: Close this connection
    #: Reset the RBRquartz3 BPR network interface
    ?: Print this menu
  4. Enter “T” to enable the transparent mode.

  5. Enter “clock” to reset the instrument clock if necessary.

  6. Enter deployment parameters, for example:

    deployment starttime = 20000101000000, endtime = 20991231235959
    sampling mode = continuous, period = 63
  7. Erase the memory and start the deployment.

    enable erasememory = true

For more information on deploying the RBRquartz³ BPR with Ethernet, see Ethernet commands.