RBRcoda³ T.ODO

The RBRlegato³ can integrate the RBRcoda³ T.ODO sensors. The measurement principle of optical dissolved oxygen sensors is based on the luminescence quenching by oxygen.

rbrcoda3 t odo angled 2
Fig. 1.    RBRcoda³ T.ODO
Parameter Value

Measurement range

0-1000μmol/L

Calibrated range
Concentration
Saturation
Temperature

 
0-500μmol/L
0-120%
1.5 to 30°C

Initial accuracy
For | fast
For standard, | slow

 
Maximum of ±8μmol/L or ±5%
Maximum of ±2μmol/L or ±1.5%

Resolution
For | fast
For standard
For | slow

 
<1μmol/L (saturation 0.4%)
<0.5µmol/L (saturation 0.2%)
<0.1μmol/L (saturation 0.04%)

Time constant

<1s | fast, <8s standard, <30s | slow

Optical dissolved oxygen measurements require pressure correction for highest accuracy. When installed on an instrument with a pressure sensor, this correction is done automatically.

Store the sensor in the dedicated storage cap, included with the instrument. Rehydrate for five days before deployment. See RBRcoda³ T.ODO care and maintenance for more information.