Quick start guide
The RBRquartz³ Q | plus uses an integrated Paroscientific Digiquartz® pressure sensor. It offers a flexible measurement schedule, standard and burst sampling, large memory, and more power for extended deployments. The RBRquartz³ Q | plus offers a variety of interfaces, including an internal USB-C port and an optional external port, which can be configured as USB, RS-232, or RS-485.
Included with your instrument
-
USB-C desktop cable with adaptor
-
O-rings, silicone compound, O-ring removal tool, and reusable desiccant
-
Refill adaptor, syringe with stopper, dispensing needle, syringe kickstand, and buffer fluid
-
External pressure adaptor
-
Cap screws and hex keys
-
USB stick containing Ruskin software and documentation
-
Calibration certificates*
*Calibration certificates are available in Ruskin. To download, connect your instrument, go to Information, and select the Download button.
Ruskin software
Find Ruskin software for Mac and PC on the USB stick included with your instrument, or visit rbr-global.com/products/software.
Deploy
-
Open the instrument.
a. b. c. d.
-
Connect the USB-C desktop cable to the port located inside the instrument body.
USB-C port
-
The instrument will appear on Ruskin.
-
Select the required Sampling mode and speed.
-
Select UTC or Local to synchronise the instrument’s clock to the computer.
-
Choose whether to start Now or enter your desired Start date/time.
-
Review the estimated End date to ensure it fulfils the deployment requirements. Longer deployments can be achieved with better battery chemistry or lower sampling speeds.
-
Click Enable to start the deployment.
Download
-
Open and remove the battery end-cap.
-
Connect to the instrument.
-
Select the Download... button and choose a location to store the dataset.
Deployment checklist
RBR ships instruments with new lithium thionyl chloride batteries and fresh desiccant capsules included, and the O-ring installed, so that the instrument is ready for its first deployment.
For any subsequent deployment:
-
Install new batteries.
-
Install fresh desiccant (orange).
-
Inspect, clean, replace, and lubricate the O-ring.
| Always remove the batteries from the instrument during long-term storage! Doing so will prevent internal damage due to battery leakage and/or corrosion. |