Quick start guide
RBR standard Instruments can measure up to 10 parameters at rates up to 32Hz. They have a rugged design, long deployment life, and are intuitive to use.
Included with your instrument
-
USB-C desktop cable with adaptor
-
O-rings, silicone compound, O-ring removal tool, and reusable desiccant
-
USB stick containing Ruskin software and documentation
-
Calibration certificates*
*Calibration certificates are available in Ruskin. To download, connect your instrument, go to Calibration, and select the Download button.
Ruskin software
Find Ruskin software for Mac and PC on the USB stick included with your instrument, or visit rbr-global.com/products/software.
Support
To access support within the Ruskin app, navigate to Help > Comment on Ruskin....
For technical support, please reach out to [email protected], call +1 613 599 8900 (UTC-5), or visit rbr-global.com/support/service.
Deploy
| RBR ships all instruments preconfigured to sample at the fastest available rate, with twist activation enabled, and the battery end-cap set to PAUSE. |
-
Remove any sensor protection caps if fitted
-
Turn the battery end-cap to the RUN
position.
The instrument will vibrate with one long pulse to indicate it has started sampling. -
Deploy the instrument.
-
To pause logging, turn the battery end-cap to the PAUSE
position.
The instrument will vibrate with three short pulses to indicate it has stopped sampling.
Alternatively, you can configure your instrument to your required sampling regime before enabling and deploying.
Download
-
Fully unscrew and remove the battery end-cap.
-
Connect the USB-C desktop cable to the USB-C port. The instrument will appear on Ruskin.
-
Click Download... and choose a location to store the dataset.
Deployment checklist
RBR ships all instruments with new, highest capacity batteries and fresh desiccant capsules included, and the O-ring installed, so that the instrument is ready for its first deployment.
For any subsequent deployment:
-
Install new batteries.*
-
Install fresh desiccant (orange).
-
Inspect, clean, replace if necessary, and lubricate the O-ring.
*All batteries must have identical chemistry and manufacturer as each other.
Fresh desiccant Exhausted desiccant
| Always remove the batteries from the instrument during long-term storage! Doing so will prevent internal damage due to battery leakage and/or corrosion. |