Example 24: corr_irr2 - generic irradiance and PAR
This equation is a refined and simplified version of the one given in Example 23, and will typically be used to present irradiance in μW/cm2/nm, or PAR in μmol/m²/s.
Then applies a linear equation to convert to final units:
where
-
c0 is a user-adjustable dark offset value.
-
c1 is the primary slope coefficient which transforms the value to final units.
-
c2 is an immersion factor, which corrects for use in a medium other than was used for calibration; water compared with air, for example.
-
R is the raw voltage ratio measured by the logger.
-
x0, x1, x2 are factory-determined coefficients of a quadratic term, used to correct the sensor’s dark current for temperature effects.
-
x3 is the temperature at which the optical calibration was performed.
-
Tsensor is the internally measured temperature of the sensor in °C.
-
Irrtempcompensated is the final value of irradiance or PAR, temperature compensated and presented in final calibration units.
Examples
>> channel 3 type
<< channel 3 type = irr_05
Confirm the channel type.
>> calibration 3 datetime = 20210901085500, c0 = 0.000, c1 = 5079.0, c2 = 1.2
Set the core coefficients.
>> calibration 3
<< calibration 3 label = irradiance_00, datetime = 20210901085500, c0 = 0.000, c1 = 5079.0, c2 = 1.2, x0 = 0.098, x1 = 0.023, x2 = 0.00037, x3 = 23.4, n0 = 4
Request confirmation of all calibration coefficients.