Skip to main content
Skip table of contents

Retrieving data

Tools needed

In order to retrieve data from the RBRcervello you need to use the diagnostic menu.

  1. Make sure the unit is powered.
  2. Connect the provided USB patch cable to the instrument and computer.

    Make sure that Ruskin isn't running when you connect the instrument.

  3. Open a serial terminal as described in Using the USB port.
  4. Disable the instrument using the "disable" command.
  5. Check that all the instruments are disabled using the "status" command.
  6. Use the "mount-storage" command to expose the data to the host computer.

    TEXT
    RBRcervello> mount-storage 
    Checking deployment status...
    Stopping services...
    Mounting data partition as read-only

    In MacOS the USB cable needs to be unplugged and replugged to the laptop in order for the drive to be detected.

  7. Copy the data from the "RBRcervello" drive that is shown on the computer.
  8. Use the "unmount-storage" command to cleanly disconnect the drive.

Drives after using mount-storage

JavaScript errors detected

Please note, these errors can depend on your browser setup.

If this problem persists, please contact our support.