O2 - compensated for pressure and salinity
Let us consider the RBRconcerto³ C.T.D.ODO, where the ODO channel is an integrated RBRcoda3 T.ODO. The RBRcoda3 T.ODO transfers both the foil temperature and the dissolved oxygen concentration (not compensated for salinity). The RBRconcerto³ calculates the concentration compensated for salinity and pressure first:

where
O2corr is the corrected O2 concentration, compensated for salinity and pressure,
O2unc is the uncompensated O2 concentration returned by the RBRcoda3 T.ODO,
C0 and C1 are corrections and scaling factors for the uncompensated O2 concentration,
C2 is a correction factor for pressure,
S is the salinity in PSU,
P is atmospheric pressure in dbar,
and

where T is the water temperature (in °C), and

which correspond to Garcia and Gordon coefficients.