Can I get a calibration range beyond +35°C?
Yes. We can perform an extended calibration in 5°C steps up to +50°C. Please note that the standard accuracy of ±0.002°C from -5°C to +35°C changes slightly to ±0.003°C for the extended range from +35° to +50°C.