Skip to main content
Skip table of contents

2.8.0

Release date: June 13 2019

Notice

This notice pertains to the following instruments:

RBRvirtuoso3, RBRduo3, RBRconcerto3, RBRmaestro3, RBRquartz3, RBRbrevio3, RBRsolo3, and RBRduet3

The notice pertains to the following sampling regimes:

Averaging, Bursting, Tide, or Wave

Recently it was discovered that Generation3 instruments using an intermittent sampling mode (bursting, averaging, tides, and waves) would present incorrect deployment autonomy estimates - the estimates are shorter than actual behaviour. The cause of the discrepancy is due to an incorrect sleep value used in the autonomy engine. Deployments where sampling occurs only for a small fraction of the burst interval are affected most.

Please contact support@rbr-global.com  for any questions or concerns regarding this notice.

New Features

  • Added support for the RBRglissando
  • Added fw 1.050 (200)

Improvements

  • RSK-6268 - Decimal places reported on calibration certificates have been reduced
  • RSK-6288 - Mouse cursor changes when in annotation mode
  • RSK-6635 - Simulated deployments life doesn't match physical instrument due to memory size estimation
  • RSK-6642 - Fixed potential deadlock in Development kit configuration
  • RSK-6646 - Display channel error code in data table
  • RSK-6648 - Added calibration stage to the channel calibration object
  • RSK-6665 - Pulled generation 3 part code descriptions into configuration files
  • RSK-6694 - Restrict instrument derived channels when fast32 feature is included on instrument
  • RSK-6700 - Decreased time required to reprocess tide and wave data
  • RSK-6701 - Reprocessing data no longer blocks the UI
  • RSK-6708 - Fixed increased CPU usage caused by UDP server
  • RSK-6710 - Fixed potential deadlock issue on the UI when closing Ruskin
  • RSK-6713 - Added default instrument settings to simulated instruments to display on the parameters tab
  • RSK-6726 - Fixed slow transition through time breaks in plot
  • RSK-6730 - Opening RSK file now sets focus in the Navigator

Bug Fixes

  • RSK-6538 - Simulated RBRduet T.ODO displays incorrect sampling rates
  • RSK-6719 - Ruskin is trying to read streaming data from a Wi-Fi port
  • RSK-6720 - Disposing of the instrument editors throws exceptions
  • RSK-6722 - Ruskin left a hanging application in the background
  • RSK-6723 - RBRquartz Q live plot is not being created
  • RSK-6731 - Reprocessing tide data doesn't work
JavaScript errors detected

Please note, these errors can depend on your browser setup.

If this problem persists, please contact our support.