2.24.1 - Internal
Release date: September 23, 2025
New Features
Added support for the Pro-Oceanus Mini CO2 sensor
Calibration certificate downloads, through Ruskin, now include attached sensors
Added support for firmware v1.160 (104) for RBRvirtuoso3, RBRduo3, RBRconcerto3, RBRmaestro3, RBRbrevio3, RBRquartz3
Added support for firmware v2.2.1 (120) for RBRlegato4
Added support for firmware v1.005 (204) for RBRcoda Tu
Added support for firmware v1.008 for FE-serial3 modules
Added RBR Instrument Guide (RIG) for 2025 RBRfermata
Added RBR Instrument Guide (RIG) for RBRlegato4
Updated RBR Instrument Guide (RIG) for RBRtridente and RBRquadrante sensors
Added production test configuration for Gen 3.5 products (PTEST_1Hz, PTEST_MAX)
Improvements
RSK-12301 - Updated default coefficients for 380nm Irradiance channels
RSK-12456 - Added 380nm SRI to radiometer calibration certificate
RSK-12486 - Warn users when deploying small instruments, which do not support Lithium thionyl chloride batteries, having a power reading >2V
RSK-12527 - More graceful handling of RBR instruments which are unsupported on the current version of Ruskin
RSK-12586 - Certificate headers now accommodate longer instrument names
Bug fixes
RSK-12295 - Unable to load RSK files with a large amount of events/errors
RSK-12375 - Unable to load RSK files with a large amount of events/errors
RSK-12412 - Maps are not longer loading for profile navigation
RSK-12473 - ODO temperature calibration results are time shifted
RSK-12484 - Unable to enable instrument in Averaging mode
RSK-12494 - RBRcoda Tu simulations provide ranging options
RSK-12526 - Verification of RBRcoda Tu calibration fails on opt_29 channels
RSK-12553 - Factory configuration fails due to lock engaging before process completes
RSK-12560 - Mensor model is incorrect on the pressure calibration certificate for 410018YM
RSK-12594 - Downsample caches are not being recalculated for derived channels
RSK-12601 - MacOS hidden files incorrectly parsed as firmware images
RSK-12617 - RBRcoda T.ODO linear calibration displays raw values in the calibrated column
RSK-12621 - Paroscientific Barometer sensor serial numbers are not recognized by Ruskin
RSK-12628 - Added support for developer firmware versions for small Gen3 instruments
RSK-12644 - RBRcoda Tu shipping personality record isn't being generated
RSK-12647 - Tcal download failed to generate a file for legacy instruments
RSK-12649 - Sampling parameter validation between Ruskin and the RBRduet3 T.ODO, do not match