2.20.9
New Features
- Added support for firmware v1.150 (104) for RBRvirtuoso3, RBRduo3, RBRconcerto3, RBRmaestro3, RBRbrevio3, RBRquartz3
- Updated Quick Start Guide (QSG) for RBRquartz3 BPR|zero
- Updated RBR Instrument Guide (RIG) for RBRquartz3 BPR|zero
- Updated RBR Instrument Guide (RIG) for RBR CT and CTD instruments
- Updated RBR Instrument Guide (RIG) for Sensors, Compact instruments, and Standard instruments
- Updated RBR Instrument Guide (RIG) for RBRlegato3
- Added factory support for TRI and QUAD part numbers
- Factory configures all instruments via configuration files. If not present will then look at tokenizing part number
- Added support for STID16, STRID17, and STRID18 sensor packs
Improvements
- RSK-10251 - Updated default pressure compensation coefficient for ODO concentration
- RSK-10542 - Improved optical PAR calibrations with hardcoded C0 coefficients
- RSK-10684 - We now handle normalizing the quadrante data in the downloaded RSK file
RSK-10725 - Updated Gordon and Garcia solubility constants
- RSK-10746 - Temperature references are ordered coldest to warmest on certificates
- RSK-10767 - Added lisocl2 battery option for RBRduo3 T.T
- RSK-10801 - Updated Ccal resistance reference values
- RSK-10837 - Added 750dbar option to legacy pressure calibration routine
- RSK-10842 - Updated settling/read times for RBRcoda T.ODO angled fast foil
- RSK-10868 - Added channel calibration range parameter to differentiate it from maximum range
- RSK-10894 - Allow Transmissometer channel to be user calibrated
- RSK-10904 - Added support for more RBRtridente channel types
- RSK-10907 - Updated Pcal conductivity stability threshold
Bug fixes
- RSK-10617 - Fixed NPE on generic RBRtridente gain calibration process
- RSK-10625 - Fixed calibration equation sizing on Chlorophyl and Fluorometer certificates
- RSK-10635 - Autocommit was missing on some SQL calls preventing recalculation of samples following coefficient updates
- RSK-10651 - Fixed lock monitor so we can update firmware on FE cards
- RSK-10672 - Firmware upgrade dialog shows a locked null version on virgin tridente
- RSK-10737 - Fixed an offset error when downloading binary chunks using variable chunk sizes
- RSK-10753 - Fixed NPE on Tcal batch calibration process
- RSK-10891 - Configuration table for Wetlabs ECO puck sensor on FE-serial3 had some incorrect settings which prevented communication
- RSK-10928 - Some sensors can display 'infinity' for sleep power under some sampling conditions