2.20.10 - internal
Release date: May 28, 2024
New Features
Added support for firmware v1.153 (104) for RBRvirtuoso3, RBRduo3, RBRconcerto3, RBRmaestro3, RBRbrevio3, RBRquartz3
Added support for firmware v1.064 (107) for RBRsolo3, RBRduet3, RBRcoda3
Added support for firmware v1.070 (200) for RBRcoda3 T.ODO
Added support for updated RBRcoda3 T.ODO|slow sensor
Added support for updated electrode pH sensor
Added support for calibration of the RBRcoda Tu sensor
Improvements
RSK-10326 - Updated memory usage display on RBRcoda instruments to show memory used, rather than memory disabled, when logging
RSK-10961 - No longer generate calibration certificate from the RBRtridente gain calibration process
RSK-10976 - Updated the resistance values for Conductivity calibration
RSK-11020 - Added Arena part number validation when loading JSON configuration file
RSK-11022 - Removed the pressure sensor serial number from the pressure calibration certificate
RSK-11071 - Added support for RBRtridente STRID111
RSK-11119 - We now hard code C0 coefficient to 0 for quadrante calibrations
Bug fixes
RSK-10870 - Fixed a display issue where the version of an incompatible RSK file was showing as 0.0.0
RSK-10922 - Fixed issue upgrading old RSK schema 1.9.0 file
RSK-10944 - Fixed a factory configuration issue with default temperature setting during 'Configure for shipment'
RSK-10971 - Fixed an RBRtridente calibration issue where plateaus were being applied to the wrong channel
RSK-10988 - Improved error condition recovery during RBRtridente firmware updates
RSK-11047 - Fixed Comment on Ruskin when submitting a QA ticket
RSK-11072 - Warning message to update Ruskin when detecting an unknown firmware version has been improved to show at the appropriate times
RSK-11138 - Reset events are now properly pushing the clock to 00:00:00 of the following day
RSK-11142 - Fixed voltage ratio gain correction for RBRquadrante downloads