Hardware overview

Required tools

  • 5mm Hex/Allen key.
  • Nitrile gloves.


Opening the RBRcervata

The RBRcervata is serviceable by opening the "battery end-cap". The battery end-cap will be the side with the connectors on it. Only the battery end cap needs to be opened to access the controller, desiccant and battery.

  • Remove the two opening screws from the end with connectors using a 5mm Hex/Allen key.


  • Pull carefully, but do not twist.

Inspecting O-rings

Pay attention to the following areas:

  • The surface of the O-ring itself
  • The mating surface on the inside of the case between the threads and the open end
  • The inner surfaces of the groove in the end cap where the O-ring sits

Any dirt present should be removed by wiping thoroughly with a soft, lint-free cloth. When cleaning, never use any material or tool which could scratch the O-ring or any of its mating surfaces. If any dirt is present in the O-ring groove, then the O-ring should be removed as described below to allow thorough cleaning of the groove. If an O-ring needs to be removed for any reason, it should be replaced.

If the O-ring is scratched, cut, distorted, perished or defective in any other way it must be replaced. If any of the surfaces of the O-ring groove are scratched, pitted or otherwise damaged, the logger may need to be returned to RBR for refurbishment: please contact us for advice.

Replacing an O-ring

  1. Lever the O-ring from its groove. Use a soft plastic or wood tool; do not use a metal screwdriver or any other tool which may scratch the surfaces of the O-ring groove, doing so will probably render the end cap useless. Slide the O-ring out of its groove and off the logger. The O-ring may need to stretch quite a bit as it is pushed off; this requires some effort but can be done by hand.
  2. Thoroughly clean the O-ring groove, taking great care not to scratch or damage it in any way. Inspect it carefully, and contact us for advice if it appears damaged. Apply a light film of silicone grease to all three inside surfaces of the groove, being careful not to trap any dirt, hairs or lint.
  3. Select the proper O-ring, see part numbers below, and make sure it is not damaged. Lubricate it with a very light film of silicone grease to ease its installation.
  4. Install the new O-ring by pushing it into place and popping it into its groove.
  5. When the new O-ring is in place, inspect it once more for scratches and dirt, and wipe away any silicone grease deposited on the end cap.

Replacing the desiccant canister

Replacing the desiccants is recommended before each deployment. The use of fresh desiccants will keep the RBRcervata compartment dry and prevent the electronics from being damaged due to condensation inside the controller. As a preventative measure, RBR recommends always use fresh desiccant packs and, when possible, service the instrument in a cool, dry place.

  • Remove the previous canister by pressing the black release mechanism and sliding the canister as shown below.


  • Insert the new canister by pressing the release mechanism with the canister itself and slide it in until it clicks into place.

Inserting/removing the micro SD card

Please ensure that the RBRcervata is fully powered down and is not connected to internal or external power. The SD card is held by a metal carrier that keeps it in the correct position.

  • Slide the metal carrier on top of the micro SD card to the "Unlock" position, as stated in its engravings. The SD card should pop up as shown below.

  • Grab the SD card by carefully sliding it away from the metal carrier.

  • After retrieving the data from it, insert it back by sliding it into the metal carrier. Do not put the SD card directly over the contacts, it should be installed in the metal carrier before locking it.

  • Lower the metal carrier with the SD card in it, and slide it into the "Lock" position.

Installing the battery pack

  • Remove battery top foam.

  • Insert the battery pack and add battery top foam passing the battery cable through it as shown below.


  • Connect the battery to the electronics and check for LEDs.


  • Insert the battery end-cap making sure the battery cable is not interfering and using the pin on the end cap for alignment.


  • Tighten the opening screws to 1.7Nm.


Opening screws

Torqued to 1.7Nm

O-rings


SD card holder


Alignment pin