Consider an RBRconcerto³ C.T.D.DO logger, where the DO channel is a RBRoptode. The RBRoptode transfers both the foil temperature and the dissolved oxygen concentration (not compensated for salinity). The RBRconcerto³ calculates the concentration compensated for salinity first, then the air saturation (in %) via the following equation: where

Solubility is calculated via Gordon and Garcia as: where S is the salinity in PSU and Ts is defined as: with T being the water temperature in °C and: Air vapour pressure (in dbar) is calculated as: The corresponding logger coefficients are:


>> calibration 9 type
<< calibration 9 type = doxy22

Confirm the channel type.

>> calibration 9
<< calibration 9 type = doxy22, datetime = 20171201000000, n0 = 4, n1 = 5, n2 = 8, n3 = value

Request confirmation of all channel indices.