Full, in-situ derivation of salinity requires that conductivity, temperature and pressure are all measured, so a simple RBRconcerto C.T.D will be used as an example.

The equation relating salinity to the three underlying parameters is the Practical Salinity Scale of 1978, often referred to as PSS78: for further information refer to the section Practical Salinity of Seawater. The equation involves a rather fearsome looking series of polynomials combined in various ways: mercifully the coefficients are all empirically determined constants, and all values are embedded in the logger.

Salinity is a 'pure' derived parameter which has its own channel assigned to it, but there is no underlying measurement hardware for salinity itself; it simply uses the outputs of the conductivity, temperature and pressure channels. This makes its specification rather sparse: there are no coefficients in either of the 'c' or 'x' groups; all that is needed is to specify the indices in the 'n' group.

Because hydrostatic pressure is used in the salinity equation, it also accommodates the presence of a channel to measure atmospheric pressure. In practice most loggers, like the RBRconcerto C.T.D in this example, will not have an "atmospheric pressure" channel, so the 'n' group index will be set to the text "value", and the logger will use the substitute value specified by the "settings atmosphere" command.

In our example:

  • n0 is the index of the temperature channel, 3 in this example,
  • n1 is the index of the pressure channel, 2 in this example,
  • n2 is the index of the conductivity channel, 1 in this example,
  • n3 is the index of the atmospheric pressure channel; not present in this example, so set to "value".

 

Example commands:

>> calibration 4 type

<< calibration 4 type = sal_00

( Confirm the channel type )

>> calibration 4

<< calibration 4 type = sal_00, datetime = 20130401120013, n0 = 3, n1 = 2, n2 = 1, n3 = value

( Request confirmation of everything )

 It is not uncommon to monitor salinity using a logger with only conductivity and temperature (C.T) channels, deployed at a constant depth. In this case we might have:

>> calibration 4

<< calibration 4 type = sal_00, datetime = 20130401120013, n0 = 2, n1 = value, n2 = 1, n3 = value

( Request confirmation of everything ) 