valvesegments
Usage
>> valvesegments [ count ]
Security
Unsafe.
Description
This is a specialized command available in firmware versions 1.142 or later. It is used in BPR|zero logger systems to help implement a more flexible schedule of valve episodes than was previously available.
If desired, the schedule of valve episodes can now be divided into a series of segments. Each segment effectively has its own "start time", and an interval between episodes that can be specified independently of other segments. This command sets or reports the number of segments that will be used during the schedule.
count [= 1...4] reports or sets the number of segments in the valve schedule. The minimum number of segments is 1, and the maximum is 4.
Refer to the valvesegment command for information on accessing the parameters for each segment.
Examples
>> valvesegments
<< valvesegments count = 2
>> valvesegments count = 3
<< valvesegments count = 3
Errors
Error E0114 feature not supported by hardware
An attempt was made to use the command with a logger that does not have the BPR|zero valve controller installed.
Error E0105 command prohibited while logging
Settings may not be modified while logging is in progress; reading them is permitted.
Error E0108 invalid argument to command
The argument supplied is out of range, or is not recognized as a valid argument.
Error E0111 command failed
An unidentified problem prevented the operation from being completed.
Error E0702 no devices configured
The configuration of the valve controller interface on this instrument is not correct.
Error E0703 device schedule inconsistent
Parameters in one of the enabled segments are not consistent.