Skip to main content
Skip table of contents

Example 15: deri_bprpres and deri_bprtemp, BPR channels

Loggers with BPR channels interface a Paroscientific, Inc. transducer. The logger measure precisely the output frequencies from the transducer. Those transducers generally outputs two signals, one for pressure and one for temperature.

There are two channels for the periods measured (peri00 and peri01). But in order to convert them in a meaningful pressure and temperature, the logger provides two channels which implement the calibration equation from Paroscientific, Inc. They rely on two type of equations: deri_bprpres (pressure in dbar) and deri_bprtemp (temperature in °C).

deri_bprpres equation


  • The coefficients x0 ... x10 are provided by Paroscientic, Inc. and are mapped according to the following bijection:

Paroscientific Inc. coefficient

U0

C1

C2

C3

D1

D2

T1

T2

T3

T4

T5

deri_bprpres coefficient

x0

x1

x2

x3

x4

x5

x6

x7

x8

x9

x10

  • n0 is the index of the pressure period channel, in this example 1

  • value(n0) is the final output value of the pressure period channel in picoseconds.

  • n1 is the index of the temperature period channel, in this example 2

  • value(n1) is the final output value of the temperature period channel in picoseconds.

deri_bprtemp equation


  • The coefficients x0 ... x3 are provided by Paroscientic, Inc. and are mapped according to the following bijection:

Paroscientific Inc. coefficient

U0

Y1

Y2

Y3

deri_bprtemp coefficient

x0

x1

x2

x3

  • n0 is the index of the temperature period channel, in this example 2

  • value(n0) is the final output value of the temperature period channel in picoseconds.

Examples

CODE
>> calibration 3 type
<< calibration 3 type = bpr_08
>> calibration 4 type
<< calibration 4 type = bpr_09

Confirm the channel type.

CODE
>> calibration 3 datetime= 20171123120721, x0 = 5.8310298E+00, x1 = -24.514029E+03, x2 = -573.64117E+00, x3 = 76.129281E+03, x4 = 35.688001E-03, x5 = 0.0000000E+00, x6 = 30.413169E+00, x7 = 664.14898E-03
<< calibration 3 datetime = 20171123120721, x0 = 5.8310300e+000, x1 = -24.514030e+003, x2 = -573.64115e+000, x3 = 76.129280e+003, x4 = 35.688000e-003, x5 = 0.0000000e+000, x6 = 30.413170e+000, x7 = 664.14899e-003
>> calibration 3 datetime = 20171123120721, x8 = 58.803409E+00, x9 = 180.91160E+00, x10 = 0.0000000E+00
<< calibration 3 datetime = 20171123120721, x8 = 58.803408e+000, x9 = 180.91160e+000, x10 = 0.0000000e+000

Set the pressure coefficients provided by Paroscientific, Inc.

CODE
>> calibration 4 datetime = 20171123120722, x0 = 5.8310298E+00, x1 = -3.8981210E+03, x2 = -10.493120E+03, x3 = 0.0000000E+00
<< calibration 4 datetime = 20171123120722, x0 = 5.8310300e+000, x1 = -3.8981210e+003, x2 = -10.493120e+003, x3 = 0.0000000e+000

Set the temperature coefficients provided by Paroscientific, Inc.

CODE
>> calibration 3
<< calibration 3 type = bpr_08, datetime = 20171123120721, x0 = 5.8310300e+000, x1 = -24.514030e+003, x2 = -573.64115e+000, x3 = 76.129280e+003, x4 = 35.688000e-003, x5 = 0.0000000e+000, x6 = 30.413170e+000, x7 = 664.14899e-003, x8 = 58.803408e+000, x9 = 180.91160e+000, x10 = 0.0000000e+000, n0 = 1, n1 = 2
>> calibration 4
<< calibration 4 type = bpr_09, datetime = 20171123120722, x0 = 5.8310300e+000, x1 = -3.8981210e+003, x2 = -10.493120e+003, x3 = 0.0000000e+000, n0 = 2

Request confirmation of all pressure and temperature calibration coefficients.

JavaScript errors detected