Skip to main content

Enabling continuous sampling

Start sampling immediately using the default configuration built into the instrument.

Assuming the default configuration is as:

TEXT
>> group create gr_pts
<< group create gr_pts
>> group gr_pts channellist=pressure_00|salinity_00|temperature_00
<< group gr_pts channellist=pressure_00|salinity_00|temperature_00

>> schedule create sch_asc_pts
<< schedule create sch_asc_pts
>> schedule sch_asc_pts grouplist=gr_pts mode=continuous
<< schedule sch_asc_pts grouplist=gr_pts mode=continuous

>> schedule sch_asc_pts grouplist=gr_pts stream=off storage=on mode=continuous period=1000 castdetection=off
<< schedule sch_asc_pts grouplist=gr_pts stream=off storage=on mode=continuous period=1000 castdetection=off

>> config create default_config
<< config create default_config
>> config default_config schedulelist=sch_asc_pts
<< config default_config schedulelist=sch_asc_pts

Ensure no time gating on deployment:

TEXT
>> deployment gate=none
<< deployment gate=none

Verify deployment:

TEXT
>> verify config=default_config dataset=dataset_01
<< verify config=default_config dataset=dataset_01 storagemode=normal state=enabled

Enable deployment:

TEXT
>> enable config=default_config dataset=dataset_01
<< enable config=default_config dataset=dataset_01 storagemode=normal state=enabled

Check current deployment status:

TEXT
>> deployment status
<< deployment status=sampling
JavaScript errors detected

Please note, these errors can depend on your browser setup.

If this problem persists, please contact our support.