Parameters you'll likely want to change are shown in bold. Calculated/output values are shown in italics. Constant/measured values unlikely to change are shown in grey.
RBRcervello RBRcervata
External battery pack None 8s7p Alkaline 4s14p LiSOCl2
Battery capacity: Wh
Instrument's baudrate: 19200 bits per second 115200 bits per second
Number of instruments:
Deep sleep duration: Seconds
Instrument deployment lifetime is calculated separately in Ruskin. However, in order to calculate the modem duty cycle (a prerequisite for estimating its power consumption), we must know some basic information about the instrument to know how much data it will produce.
Instrument deployment lifetime is calculated separately in Ruskin. However, in order to calculate the download time (a prerequisite for estimating its power consumption), we must know some basic information about the instrument with the most channels/sample rate to know how much data it will produce.
Sample rate: Hz
Number of channels:
Average power consumption: W
Direction-dependent sample reduction: %
Efficacy of direction-dependent sampling depends on configuration and deployment conditions. See the Logger3 command reference for details.
Data rate: bytes/second
Whether a HEM is included in the system, and its means of attachment to the inductive loop, have a significant impact on overall power consumption. SSM power consumption is not considered, as it draws power from the same source as the instrument, not the data controller.
Instrument attachment: No MLM MLM with direct connection MLM with ferrite holder
Interval: seconds
Activity duration: s; power consumption: W
The Airmar 200WX measures tilt, air temperature, wind speed and direction.
Enabled:
Weather number of channels: 6 9
Sampling burst speed: 1 10 Hz
Sampling burst duration: s
Sampling burst period: s
Telemetry compression reduction: %
The default reduction value – 80% – is based on empirical data collected across multiple deployments. It encompasses the effects of not only compression, but also real-world telemetry overhead from the messaging protocol (connection metadata, diagnostic data, message delimiters) and any repeated data transmitted due to telemetry drops/outages. Strictly speaking, the actual compression (quantization + differential encoding) is significantly more effective, but to show only the compression ratio would be a misrepresentation of the true amount of data transferred.
For GSM, use “-5%”: no compression is used, but there is still protocol overhead which increases the payload size.
Data rate: bytes/telemetry session
Modem: GSM Iridium RUDICS
Configuration/dialing activity duration: s; power consumption: W
Dialed, transmitting + receiving power consumption: W
Duty cycle: %
Deployment life: days